Google+ MRI Room Oxygen Monitor - CSP Medical Store
  Loading... Please wait...

Categories

  • Image 1
  • Image 2
  • Image 3

MRI Room Oxygen Monitor

Brand:
Request Quote


Product Description

Air Check O2 Sample Draw Monitor for MRI

Our Air Check O2 sample draw monitor is a specialized version of our oxygen deficiency monitor perfect for situations where the monitor cannot physically be in the concerning area. The monitor can reside up to 100ft away from the area to be tested and uses a polyurethane or equivalent 1/4 inch tube to draw the sample. For a sterile environment, this monitor is the solution. This has a 10+ year zirconium oxide sensor.

Monitor comes with a 110/220V plug-in power supply (part #L 45101).  This monitor is designed to mount in the controller's room and take a sample from the MRI Room.

Features:

  • 10 + Years No Calibration Sensor
  • Two-Alarm Relays for 19.5% and 18.0%
  • No Maintenance Required
  • 3 Year Warranty
  • Digital Display, 4-20mA Analog Output
  • No drift due to thunderstorms or barometric pressure changes
  • Wall mounting brackets and optional plug-in wall power supply
  • Built-in Flow Sample Pump
  • UL, C UL and CE Approvals

Optional Items:

  • 8 Channel Controller
  • Horn and Strobe
  • Strobe and Display

Long Last Zirconium Sensor

As a leading manufacturer of oxygen monitors, PureAire Monitoring Systems puts performance, reliability, and affordability first. Our oxygen monitors last for 10+ years with no replacement, no maintenance, and no hidden costs.

These monitors work well anywhere gas leaks would cause oxygen depletion, including in laboratories, cryogenics chambers, hyperbaric chambers, and freezers. Our monitors have a long-lasting zirconium oxide sensor, which delivers reliable monitoring against leaks of argon, helium, and nitrogen for 10 years. The device has its own audible and visual alarm, but it can be connected to fire alarm systems to amplify the reach.

The O2 monitor is easy to set up, easy to operate, and requires no calibration once installed. Supervised watchdog software checks that everything is working properly and notifies you of any inconsistencies on the monitor. Immune to performance changes from barometric pressure drops, humidity, and temperature, these monitors are reliable and cost-effective.

Specifications:

Sampling Method Sample Draw
Gases Detected Oxygen (O2)
Range 0 – 25 %
Accuracy ± 1 % of Full Scale
Operating Temperature -40 to +55 ºC
Display 3/4″ LCD Digital Display, Back Lit
Sensor Type Long life zirconium oxide sensor cell
Sensor Life > 10 years under Normal Conditions
Signal Outputs 4-20 mA Analog Output
Dual Level Alarm Relay Contacts
Built in Horn
Power Requirements 24 VDC
300 mA
Width 7.25 in
184.2 mm
Height 6.0 in
152.4 mm
Depth 5.0 in
127 mm
Weight 4.0 lb
2.0 kg
Enclosure1 Polycarbonate
Required Calibration None
No Zero or Span Pot Adjustment Required
Optional
Available without Horn. Contact PureAire for details.

1 Not intended for explosive atmospheres

 161-1000-itemtable-new.jpg

 