Google+ Mini Doppler Flow System - CSP Medical Store Web Statistics
  Loading... Please wait...

Categories

  • Image 1

Mini Doppler Flow System

Request Quote


Product Description

Product Overview

The Doppler Flow System tests both Doppler and B-mode ultrasound systems in a single unit.  The compact, easy to store and transport designed system combines the flow system, phantom and electronic flow controller into a single unit. 

Scanner selection, quality control testing, training and research can all be performed using this multi-faceted ultrasound tool. 

A wide range of targets and vessels are included in the unit.

Features

  • The Choice of attenuations of 0.5 or 0.7 dB/cm/MHz
  • Includes 404GS LE components with Grey Scale targets 
  • Combines low echo matrix with line reflectors and anechoic cyst targets at 2, 4 and 6 mm depths
  • Two 5mm vessels in the system adhere to FDA Doppler sensitivity recommendations. 
  • Flow controller with a range of 1 to 12.5 ml/sec
  • 5 preset pulse flow patterns

Benefits

  • Perform multiple tests with the same unit
  • Grey Scale targets permit measuring contrast temporal resolution and system linearity 
  • Measure maximum signal penetration in real time  
  • Determine flow rate, visibility for various angles, beam directionis and operating modes in real time
  • Determine similarities between B-mode and color flow images in real time
  • Preset programs allow for reproducible comparative testing

Warranty

  • Comes with 1 year Warranty
  • Extended Warranty plans are available. 

102-1880-itemtable.jpg

SPECIFICATIONS
 

Tissue Mimicking Material

Speed of Sound . . 1540 ±10 m/s

Attenuation . . . . 0.5 or 0.7 ±0.05 dB/cm/MHz

Vessels

Size. . . . . . . . . . 4 mm inside diameter

Location . . . . . . horizontal 2 cm below surface; diagonal at 35 degrees from 1.5 cm below surface

Grey Scale Targets

Contrast . . . . . . . -6, +6, & +12 dB relative

                               to background

Diameter . . . . . . 7 mm

Speed of Sound . . 1540 ±10 m/s

Attenuation . . . . 0.5 or 0.7 ±0.05 dB/cm/MHz

Electronic Flow Control System

Measurable

Velocities . . . . . . 0-174 cm/sec

Pulsatile

Flow Mode . . . . . preprogrammed for

0.25-2 seconds/pulse

Anechoic Cysts

Diameters . . . . . 1, 2, 4, and 7 mm

Speed of Sound . . 1540 ±10 m/s

Attenuation . . . . 0.05 ±0.01 dB/cm/MHz

Pin Targets

Diameter of nylon lines . . . . . 0.1 mm (0.004 in)

Vertical Spacing. . 5 mm at 1 cm to 9 cm deep

Horizontal

Spacing . . . . . . . 10 mm at 1 cm and 5 cm  deep

Blood Mimicking Fluid

Density . . . . . . . 1.03 g/cc

Speed of Sound . . 1550 ±10 m/sec

Scatter Size . . . . 4.7 micron average

Container Construction

Size. . . . . . . . . . 22.9 x 15.9 x 21.9 cm  

                           (9 x 6.25 x 8.625 in)

Weight. . . . . . . . 4.4 kg (9.5 lbs)

AC Input Voltage.....110 vac - 240 vac

                                   50-60Hz

Battery . . . . . . . LI-ion, rechargeable,

                            universal charger and 

                            plugs included


Vendors Other Products